ADVIA 120
Minimizing review rates and repeats utilizing proven flow cytometry
Performance Specifications
WBC | 0.02 to 400 x 103/µL |
RBC | 0.0 to 7.0 x 106/µL |
PLT | 5.0 to 3500 x 103/µL |
HB | 0.0 to 22.5 g/dL |
Retic | 0.2 to 24.5% |
Precision
Parameter | Mean | SD | CV |
---|---|---|---|
WBC | 7.5 | 0.20 | 2.7 |
RBC | 5.0 | 0.06 | 1.2 |
HB | 15.0 | 0.14 | 0.93 |
MCV | 90.0 | 0.7 | 0.78 |
PLT | 300.0 | 8.8 | 2.93 |
Retic% | 2.0 | 0.25 | 12.5 |
Test Selectivity / Throughput
CBC | 120 samples/hour |
CBC/DIFF | 120 samples/hour |
CBC/DIFF/Retic | 74 samples/hour |
CBC/Retic | 74 samples/hour |
Retic | 74 samples/hour |
Sample Modes / Volumes
Automatic Closed Tube | 175µL |
Manual Closed Tube | 175µL |
Manual Open Tube | 175µL |
Weight and Dimensions
Analytical Module with Autosampler: | |
---|---|
Weight | 193 kg / 425 lbs |
Height | 86 cm / 33.8 in |
Width | 141 cm / 55.5 in |
Depth | 68 cm / 26.8 in |
Analytical Module without Autosampler: | |
Weight | 163 kg / 360 lbs |
Height | 86 cm / 33.8 in |
Width | 81 cm / 31.9 in |
Depth | 68 cm / 26.8 in |
Workstation Specifications
Intel Pentium E6500 Dual Core 2.93 GHz
350W Desktop/Tower
3Gb/s 7200 RPM 16MB Cache Hard Drive
1 GB Memory Module
CD-RW
17” Color Monitor
Full-sized Keyboard (major languages)
Autosampler Option
150 Sample Capacity: | 15 Racks with 10 tube capacity |
Compatible tube types: | Vacutainer®, Hemogard™, Monovette® (Sarstedt 2.6 mL tubes require rack modification), Greiner MicroSample Tubes, BD MAP tubes |
Automatic discrimination of label codes: | Codabar Interleave 2 of 5 with and without check digit Code 39 Code128 EAN and JAN (8 and 13) |
Other Specifications
Onboard analytical and wash reagents (1850 CBC/Diff tests)
Automated Daily Maintenance – fully programmable